    摘要:

    激光诱导击穿光谱(laser induced breakdown spectroscopy, LIBS)结合门控光电倍增管(photo multiplier tube, PMT)技术中的信号采集装置是影响LIBS分析系统小型化与集成化的关键组件之一,通过采用微控制芯片与AD9226芯片设计实现了7.6 MHz采集频率的12 bit PMT信号高速采集系统,并对水溶液中铅(Pb)、铬(Cr)元素进行了定量分析。实验结果表明,LIBS结合门控PMT分析系统采用高速采集系统建立的Pb、Cr元素定标曲线拟合度分别为0.988、0.978,相对标准偏差分别为3.56%、6.57%,检出限分别为0.013 mg/L与0.087 mg/L,相比未采用PMT的LIBS检测技术降低了一个数量级。与电感耦合等离子体质谱法对比结果显示,对质量浓度为7.0 mg/L的Pb溶液与0.7 mg/L的Cr溶液,检测结果的相对误差分别为2.81%、3.26%。因此PMT信号高速采集系统在水溶液重金属的LIBS分析中具有良好的定量效果,且具有体积小、功耗与成本低的优势,有利于进一步推进LIBS技术在水质重金属检测领域中的应用。

    Abstract:

    LIBS (laser induced breakdown spectroscopy) combined with the signal acquisition device of gated PMT(photo multiplier tube) technology is one of the key components that affect the miniaturization and integration of the LIBS analysis system. Therefore, a 12 bit PMT signal high-speed acquisition system with a 7.6 MHz acquisition frequency was designed and implemented using a microcontrol chip and an AD9226 chip. The element lead(Pb) and chromium (Cr) were quantitatively analyzed. The experimental results show that LIBS combined with Gated-PMT analysis system adopts high-speed acquisition system to establish Pb, Cr element calibration curve fitting degree is 0.988, 0.978, relative standard deviation is 3.56%, 6.57%, detection limit is 0.013 mg/L and 0.087 mg/L, respectively. Compared with the LIBS detection technology without PMT, it is an order of magnitude lower. Compared with inductively coupled plasma-mass spectrometry, the relative errors of Pb solution with 7.0 mg/L and Cr solution with 0.7 mg/L were 2.81% and 3.26%, respectively. Therefore, the PMT signal high-speed acquisition system has a good quantitative effect in the LIBS analysis of heavy metals in aqueous solutions, and has the advantages of small size, low power consumption and low cost, which is conducive to further promoting the application of LIBS technology in the field of water quality heavy metal detection.

    参考文献
    相似文献
    引证文献
引用本文

施佳明, 闫子琪, 郭连波, 等. 门控PMT联用LIBS技术的高速采集系统研发及其应用[J]. 国防科技大学学报, 2025, 47(5): 34-40.